• Result re-calculation facility eliminates repeat runs
  • Soft Start and Stop of all Reagent and Sample probes
  • Vertical Obstruction Detection (VOD) system of XL 300 detects obstructions and protects the probe needles from damage
  • Separation of biohazardous waste for safe disposal
  • Weight based level sensing platform with alarms for D.I. water, cleaning solution and waste containers
  • Storage of Calibration Trace Data with Reagent Blank O.D. and Calibration O.D.

 

  • System    :    Automated, Discrete, Random Access, Patient Prioritised Chemistry Analyzer Fully Automated Random Access Clinical Chemistry Analyzer
  • Throughput    :    300/360# Photometric Tests per Hour Independent of test combination ; (360 tests/hr for China & Turkey only)
  • On-line tests    :    Maximum 50 tests
  • Programmable Parameters :Upto 99 PhotometricTests, 10 Calculation Items
  • Assay modes    :    1 -Point, 2-Point, Rate-A and Rate-B
  • Calibration    :    Linear, Non-linear Multipoint
  • Sample disk capacity    :    Outer ring : 50 positions for patient samples (routine) includes 5 positions for STATS Inner ring : 20 positions for standards / calibrators, 2 blanks & 8 controls
  • Sample disk    :    Primary tubes of 5, 7 or 10 ml & Sample cups
  • Reagent disk    :    Cooled reagent disk with 50 positions
  • Reagent pipetting    :    10-300 µl steps of 1 µl. Possibility of 1 or 2 reagent tests Sample pipetting    :    1-60 µl steps of 0.1 µl
  • Reagent bottles    :    20 ml, 50 ml & Special adapter for 5 ml  bottles
  • Reaction disk    :    60 Permanent Individual Hard Glass  Cuvettes
  • Photometry    :    Mono & Bi-chromatic measurements with Multi-wavelength diffraction grating with 12 wavelength options (340, 376, 415, 450, 480, 505, 546, 570, 600, 660, 700 & 750 nm)
  • Reading volume    :    250 µl
  • Auxiliary data storage    :    1,00,000 results Interface    :    RS 232
  • Power supply    :    110/220 ±10%, 50/60 Hz, 1.0 KVA
  • Computer (requirements)    :    Pentium IV with colour monitor, CD Drive, 256 MB RAM & Printer Ambient temperature    :    20-30°C
  • Ambient humidity    :    20-85 % non condensing
  • Water consumption    :    Max. 8 ltrs/hour
  • Dimension    :    840 (W) x 610 (D) x 1100 (H) mm
  • Weight    :    Max. 150 kgs. (Main unit with CPU)