• Continuous loading of 60 tubes with autoloader
  • Improved workflow-Throughput: 80 samples / hr
  • STAT mode for urgent sample testing
  • Mode Type : Whole Blood, Predilute, Capillary
  • Autoloader : 60 Tubes; Continuous Loading with auto-mixing
  • 31 Parameters : CBC: WBC, RBC, Hb, HCT, MCV MCH, MCHC, RDW-CV, RDW-SD, PLT, MPV, PCT, PDW-SD, PDW-CV, P-LCR, P-LCC, DIFF: Neutrophils % & #, Lymphocytes % & #, Monocytes % & #, Eosinophils % & #, Basophils % & # *ALY % & #, *LIC % & #, *NLR *Research use only
  • Principle of Measurement: RBC/PLT/WBC: Electrical Impedance WBC DIFF: Tri-angle Laser Flowcytometry Hb: Cyanide Free Colorimetry MCV: Measured HCT: Calculated Histograms : 3 (RBC/WBC/PLT) Scattergrams : 3 (2D Scattergram), 1 (3D Scattergram) QR Code Reader For reading sample bar codes & direct input of QC values
  • Sample Volume Whole Blood : 20 μL Pre-diluted : 20 μl Capillary : 20 μl CBC Mode : 13 μL
  • Linearity Range WBC : 0 - 300 (x 103/μL) RBC : 0.00 - 8.50 (x 106/μL) Hb : 0 - 25.0 (g/dL) HCT : 0 - 67 (%) PLT : 0 - 3000 (x 103/μL)
  • Calibrator ELite H Cal
  • Tri-level Controls Erba H5 CON L, N, H Open Vial Stability at 2-8˚C: 16 Days
  • Throughput 80 Tests/Hr
  • Quality Control 500 files with L-J graph X-Bar Analysis
  • Data Storage 1,00,000 Results with Graphs
  • Dimension (mm) 650x550x610
  • Weight (Kg) 59
  • Reagents ELite 580 Dil ELite H580Lyse1 ELite H580Lyse2 ELite H580Lyse3 Elite H Clean 
  • Interfaces 4 USB ports (For connecting external devices including external printer) + 1 LAN Port (HL 7, TCP/IP; LIS connectivity)
  • Operating Environment Temperature : 15-40˚C Humidity : 30-90%
  • Power Requirement A.C.100-240V with built-in voltage stabiliser 50/60Hz, 200VA